Top 10 Myths About Container Seal Testing I
Myth: Dye ingress testing, such as the ability of a container to self-seal, is a valid CCIT method
Fact: Compendial or International Standards Organization (ISO) dye ingress testing examines the ability of a multidose stopper to reclose when repeatedly punctured with an injection needle (1,2). However, these tests are not validated, reliable, or sensitive enough for overall package integrity verification (3), and results are often variable and subjective. Spectrophotometric dye detection does not eliminate the risk of false positive or negative results. Another common misconception about dye ingress is that test method sensitivity is defined as the limit of quantitation of dye detection. In contrast, the sensitivity of a leak test method is the test's ability to identify a package with a specific defect in a random mixture of leak-free and leaking containers. If the challenge conditions of the ingress test do not allow the dye to be drawn into the package, or if the product itself blocks or obstructs the leak path, then the detection method will be ineffective in identifying trace amounts of dye.
Knowledge about steel pails and other packages